The 601 Pro SeriesXL is an automated electrical-safety analyzer that meets stringent international standards for electrical-safety testing of hospital and laboratory electromedical equipment. The 601 Pro SeriesXL is also available with a built-in printer. (See โ€Models.โ€)
Standard Features
- EN60601-1, EN601010-1, and AAMI test loads (user selectable)
- Multiple enclosure-leakage points
- Multiple patient-applied-part types
- Power ON/OFF delay
- DC only for patient- and auxiliary-leakage tests
- User-programmable test sequences
- Manual, auto, step, and computer-control modes
- ASCII data transfer
- Memory for up to 1000 device-information records
|
Optional Features
| Specifications |
| Voltage |
| (single and dual lead) |
ย |
| Range: |
ย 0 to 300 V True RMS |
| Accuracy: |
ย DC - 100 Hz ยฑ 1.5 % of reading ยฑ 1 LSD |
|
| Insulation Resistance |
| Range: |
ย 0.5 to 400.0 Mฮฉ |
| Accuracy: |
ย ยฑ 5 % of reading ยฑ 2 LSD |
|
| Current Consumption |
| Range: |
ย 0 to 15 A ac True RMS |
| Accuracy: |
ย ยฑ 5 % of reading ยฑ 2 LSD |
|
| Mains on Applied Part |
| Applied Voltage: |
ย โฅ 110 % of mains voltage |
| Accuracy: |
ย ยฑ 2 % of reading ยฑ 6 ยตA |
|
| Supply Voltage |
|
ย 90 to 265 Vac, autoswitching |
|
| Protective Earth Resistance |
| Range: |
ย 0.000 to 2.999 ฮฉ |
| Accuracy: |
ย ยฑ 5 % of reading ยฑ 4 mฮฉ (1 A, 10 A, and 25 A test currents) (Refer to Operatorโ€s Manual for additional specs qualifying the effects on accuracy of variations in load inductance and phase angle.) |
|
| IEC601-1 and AAMI Leakage Currents |
| Range: |
ย 0 to 8000 ยตA True RMS |
| Accuracy: |
ย (per IEC601-1 or AAMI filter) |
|
ย - DC - 1 kHz ยฑ 1 % of reading ยฑ 1 ยตA |
|
ย - 1 to 100 kHz ยฑ 2 % of reading ยฑ 1 ยตA |
|
ย - 100 kHz to 1 MHz ยฑ 5 % of reading ยฑ 1 ยตA |
| DC-Only Frequency Response: |
ย DC - 5 Hz (approx) |
|
| IEC1010-1 Leakage Current |
| Range: |
ย 0 to 16000 ยตA True RMS (For some leakage tests, displayed range may be 0 to 9999 when using IEC1010 test load.) |
| Accuracy: |
ย (per IEC1010-1 figure A.1 filter) |
|
ย - DC - 1 kHz ยฑ 2 % of reading ยฑ 2 ยตA |
|
ย - 1 to 100 kHz ยฑ 4 % of reading ยฑ 2 ยตA |
|
ย - 100 kHz to 1 MHz ยฑ 10 % of reading ยฑ 10 ยตA |
| DC-Only Frequency Response: |
ย DC - 5 Hz (approx) |
|
| ECG Simulation and Performance Testing |
| ECG Complex: |
ย 30, 60, 120, 180, 240 BPM |
| Performance |
ย |
| Pulse: |
ย 30, 60 BPM, 63 ms pulse width |
| 600 to 700 ยตs rise and fall time |
ย |
| Sine Waves: |
ย 10, 40, 50, 60, 100 Hz |
| Square Wave: |
ย 0.125, 2.000 Hz (50 % duty cycle) |
| Triangle Wave: |
ย 2 Hz, 2 mV |
|
| Arrhythmia |
|
ย Atrial fibrillation |
|
ย Atrial flutter |
|
ย Atrial tachycardia |
|
ย Idioventricular |
|
ย PVC1 |
|
ย R on T |
|
ย Run |
|
ย Ventricular fibrillation |
|
ย Ventricular tachycardia |
| Performance: |
ย ยฑ 2 % of reading for rate or ยฑ 5 % of reading for amplitude, fixed at 1 mV peak on a Lead II ECG connection (except for triangle wave, which is 2 mV peak-to-peak) |
|
| Dimensions |
|
ย 16.62 in L x 11.75 in W x 5.56 in H |
|
ย (42.2 cm L x 30 cm W x 14.1 cm H ) |
|
| Weight |
|