ESA612 - Electrical Safety Aanlyzer  

 

  • ภาพรวม
  • ลักษณะเด่น
  • ลักษณะเฉพาะ
  • ดาวน์โหลด
The ESA612 Electrical Safety Analyzer represents the next generation in test equipment for biomedical and clinical engineering professionals that perform field service on medical equipment throughout their facilities, in clinics, and anywhere onsite service is required. Portable, lightweight, and designed for operation in tight spaces, the ESA612 offers the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool.

With selection of two test loads, this versatile product can be used worldwide for compliance testing to preventative maintenance electrical safety standards of choice: ANSI/AAMI ES1:1993 (NFPA-99), IEC62353 (VDE 751) and AN/NZS 3551

The versatility of the multifaceted ESA612 tester is further expanded with optional automation software, which speeds and simplifies testing and provides high-end-analyzer productivity at software-level investment. Ansur-automated ESA612 standardizes test procedures, compares results to standards limits, and generates and stores reports for total digital data management.
Key Features:
  • Portable, ergonomic, lightweight, and easy to use
  • Large, easy-to-read display with adjustable contrast
  • Human-factors-designed user interface
  • Tilt stand design for stand-up testing in field environments
  • Five applied parts jacks and easy ECG snap connection with optional expander box
  • ECG waveform tests and dual-lead measurements combine the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool
  • Replaceable mains fuses keep your unit in the field and out of the repair shop
  • Internal memory for 100 test records
  • 20 A at 120 V current capability
  • USB connection for use with Ansur and Data Viewer software (for memory download to PC)
  • Two-year extended warranty (no-cost, available after first-year calibration at the Fluke Biomedical Cleveland Service Center)
  • Optional Ansur automation software standardizes test procedures, compares results to standards limits, generates/stores reports and provides total digital data management
  • Rigorously tested for rugged field applications, with CE and CSA in addition to the Fluke-quality-design stamp of approval
Specifications
Voltage
Range (mains voltage)  90 V ac rms to 132 V ac rms
 180 V ac rms to 264 V ac rms
Range (accessible voltage)  0 V ac rms to 300 V ac rms
Accuracy  ± (2 % of reading + 0.2 V)
Voltage tests  Mains and point to point
Earth resistance
Modes  Two terminal
Test current  > 200 mA ac
Ranges  0 Ω to 2 Ω
Accuracy  ± (2 % of reading + 0.015 Ω)
Resistance tests  Earth resistance and point to point
Equipment current
Mode  ac rms
Range  0 A to 20 A
Accuracy  ± (5 % of reading + (2 counts or 0.2 A, whichever is greater))
Duty cycle  15 A to 20 A, 5 min. on/5 min. off
10 A to 15 A, 7 min. on/3 min. off
0 A to 10 A continuous
Leakage current
Modes  ac + dc (true rms)
ac only
dc only
 * Modes are available in all leakage tests with the exception of MAP leakages that are available only in true rms.
Patient load selection (input impedance)  AAMI ES1-1993 Fig.1 IEC 60601-1: Fig 15
Crest factor  ≤ 3
Ranges  0 µA to 199.9 μA
200 µA to 1999 μA
2 mA to 10 mA
Frequency response/ Accuracy  DC to 1 kHz: ± (1 % of reading + (1 μA or 1 LSB, whichever is greater))
1 kHz to 100 kHz: ± (2 % of reading + (1 μA or 1 LSB, whichever is greater))
1 kHz to 5 kHz (current > 1.6 mA): ± (4 % of reading + (1 µA or 1 LSD, whichever is greater))
100 kHz to 1 MHz: ± (5% of reading + (1 μA or 1 LSB, whichever is greater))
 Note: Accuracy for Isolation, MAP, Direct AP, Alternative AP, and Alternative Equipment leakage tests all ranges are + (2.5 µA or 1 LSD, whichever is greater)
Leakage tests  Ground wire (earth)
 Chassis (enclosure)
 Lead to ground (patient)
 Lead to lead (patient auxiliary)
 Lead isolation (mains on applied part)
 Direct equipment
 Direct applied part
 Alternative equipment
 Alternative applied part
 Point to point
Mains on applied part  100 % of mains
Differential leakage
Ranges  75 µ A to 199 µA
200 µA to 1999 µA
2 mA to 20 mA
Accuracy  ± (10 % of reading + (2 counts or 20 uA, whichever is greater))
Insulation resistance
Ranges  0.5 MΩ to 20 MΩ
20 MΩ to 100 MΩ
Accuracy  ± (2 % of reading + 0.2 MΩ)
 ± (7.5 % of reading + 0.2 MΩ)
Source test voltage  500 V dc
 250 V dc
Insulation resistance tests  Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible conductive part) and AP- NE (non-earthed accessible conductive part)
ECG performance waveforms
Accuracy  ± 2 %
 ± 5 % for amplitude of 2 Hz Square wave only, fixed @ 1 mV Lead II configuration
Waveforms: Rates  ECG Complex: 30 BPM, 60 BPM, 120 BPM, 180 BPM, and 240 BPM Ventricular Fibrillation
 Square wave (50% duty cycle): 0.125 Hz and 2 Hz
 Sine wave: 10 Hz, 40 Hz, 50 Hz, 60 Hz, and 100 Hz
 Triangle wave (Hz): 2
 Pulse (63 ms pulse width): 30 and 60
Mains voltage outlet
120 V ac  230 V ac
Mains voltage inlet power range
90 V ac rms to 132 V ac rms  180 V ac rms to 264 V ac rms
Maximum current
20 A  16 A
Hz
50 or 60  50 or 60
Physical case
Dimensions (WxDxH)  17.63 cm x 8.38 cm x 28.45 cm (6.94 in x 3.3 in x 11.2 in)
Weight  1.6 kg (3.5 lb)
Environmental
Operating temperature  10 ºC to 40 ºC (50 ºF to 104 ºF)
Storage temperature  -20 ºC to 60 ºC (-4 ºF to 140 ºF)
Operating humidity  10 % to 90 % non-condensing
Altitude  120 V ac mains supply voltage up to 5, 000 meters
230 V ac mains supply voltage up to 2,000 meters
Warranty
 Two-year extended warranty (no-cost, available after first-year calibration at the Fluke Biomedical Cleveland Service Center, otherwise standard one year warranty applies)