Fluke Biomedical >> Electrical Safety Analyzer >> ESA612
 |
ESA612 - Electrical Safety Aanlyzer |
|
 |
- ภาพรวม
- ลักษณะเด่น
- ลักษณะเฉพาะ
- ดาวน์โหลด
The ESA612 Electrical Safety Analyzer represents the next generation in test equipment for biomedical and clinical engineering professionals that perform field service on medical equipment throughout their facilities, in clinics, and anywhere onsite service is required. Portable, lightweight, and designed for operation in tight spaces, the ESA612 offers the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool.
With selection of two test loads, this versatile product can be used worldwide for compliance testing to preventative maintenance electrical safety standards of choice: ANSI/AAMI ES1:1993 (NFPA-99), IEC62353 (VDE 751) and AN/NZS 3551
The versatility of the multifaceted ESA612 tester is further expanded with optional automation software, which speeds and simplifies testing and provides high-end-analyzer productivity at software-level investment. Ansur-automated ESA612 standardizes test procedures, compares results to standards limits, and generates and stores reports for total digital data management.
Key Features:
- Portable, ergonomic, lightweight, and easy to use
- Large, easy-to-read display with adjustable contrast
- Human-factors-designed user interface
- Tilt stand design for stand-up testing in field environments
- Five applied parts jacks and easy ECG snap connection with optional expander box
- ECG waveform tests and dual-lead measurements combine the functionality of a simulator, multimeter and electrical-safety analyzer in a single test tool
- Replaceable mains fuses keep your unit in the field and out of the repair shop
- Internal memory for 100 test records
- 20 A at 120 V current capability
- USB connection for use with Ansur and Data Viewer software (for memory download to PC)
- Two-year extended warranty (no-cost, available after first-year calibration at the Fluke Biomedical Cleveland Service Center)
- Optional Ansur automation software standardizes test procedures, compares results to standards limits, generates/stores reports and provides total digital data management
- Rigorously tested for rugged field applications, with CE and CSA in addition to the Fluke-quality-design stamp of approval
|
| Specifications |
| Voltage |
| Range (mains voltage) |
90 V ac rms to 132 V ac rms |
|
180 V ac rms to 264 V ac rms |
| Range (accessible voltage) |
0 V ac rms to 300 V ac rms |
| Accuracy |
± (2 % of reading + 0.2 V) |
| Voltage tests |
Mains and point to point |
|
| Earth resistance |
| Modes |
Two terminal |
| Test current |
> 200 mA ac |
| Ranges |
0 Ω to 2 Ω |
| Accuracy |
± (2 % of reading + 0.015 Ω) |
| Resistance tests |
Earth resistance and point to point |
|
| Equipment current |
| Mode |
ac rms |
| Range |
0 A to 20 A |
| Accuracy |
± (5 % of reading + (2 counts or 0.2 A, whichever is greater)) |
| Duty cycle |
15 A to 20 A, 5 min. on/5 min. off
10 A to 15 A, 7 min. on/3 min. off
0 A to 10 A continuous |
|
| Leakage current |
| Modes |
ac + dc (true rms)
ac only
dc only |
|
* Modes are available in all leakage tests with the exception of MAP leakages that are available only in true rms. |
| Patient load selection (input impedance) |
AAMI ES1-1993 Fig.1 IEC 60601-1: Fig 15 |
| Crest factor |
≤ 3 |
| Ranges |
0 µA to 199.9 μA
200 µA to 1999 μA
2 mA to 10 mA |
| Frequency response/ Accuracy |
DC to 1 kHz: ± (1 % of reading + (1 μA or 1 LSB, whichever is greater))
1 kHz to 100 kHz: ± (2 % of reading + (1 μA or 1 LSB, whichever is greater))
1 kHz to 5 kHz (current > 1.6 mA): ± (4 % of reading + (1 µA or 1 LSD, whichever is greater))
100 kHz to 1 MHz: ± (5% of reading + (1 μA or 1 LSB, whichever is greater)) |
|
Note: Accuracy for Isolation, MAP, Direct AP, Alternative AP, and Alternative Equipment leakage tests all ranges are + (2.5 µA or 1 LSD, whichever is greater) |
| Leakage tests |
Ground wire (earth) |
|
Chassis (enclosure) |
|
Lead to ground (patient) |
|
Lead to lead (patient auxiliary) |
|
Lead isolation (mains on applied part) |
|
Direct equipment |
|
Direct applied part |
|
Alternative equipment |
|
Alternative applied part |
|
Point to point |
| Mains on applied part |
100 % of mains |
|
| Differential leakage |
| Ranges |
75 µ A to 199 µA
200 µA to 1999 µA
2 mA to 20 mA |
| Accuracy |
± (10 % of reading + (2 counts or 20 uA, whichever is greater)) |
|
| Insulation resistance |
| Ranges |
0.5 MΩ to 20 MΩ
20 MΩ to 100 MΩ |
| Accuracy |
± (2 % of reading + 0.2 MΩ) |
|
± (7.5 % of reading + 0.2 MΩ) |
| Source test voltage |
500 V dc |
|
250 V dc |
| Insulation resistance tests |
Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible conductive part) and AP- NE (non-earthed accessible conductive part) |
|
| ECG performance waveforms |
| Accuracy |
± 2 % |
|
± 5 % for amplitude of 2 Hz Square wave only, fixed @ 1 mV Lead II configuration |
| Waveforms: Rates |
ECG Complex: 30 BPM, 60 BPM, 120 BPM, 180 BPM, and 240 BPM Ventricular Fibrillation |
|
Square wave (50% duty cycle): 0.125 Hz and 2 Hz |
|
Sine wave: 10 Hz, 40 Hz, 50 Hz, 60 Hz, and 100 Hz |
|
Triangle wave (Hz): 2 |
|
Pulse (63 ms pulse width): 30 and 60 |
|
| Mains voltage outlet |
|
| Mains voltage inlet power range |
| 90 V ac rms to 132 V ac rms |
180 V ac rms to 264 V ac rms |
|
| Maximum current |
|
| Hz |
|
| Physical case |
| Dimensions (WxDxH) |
17.63 cm x 8.38 cm x 28.45 cm (6.94 in x 3.3 in x 11.2 in) |
| Weight |
1.6 kg (3.5 lb) |
|
| Environmental |
| Operating temperature |
10 ºC to 40 ºC (50 ºF to 104 ºF) |
| Storage temperature |
-20 ºC to 60 ºC (-4 ºF to 140 ºF) |
| Operating humidity |
10 % to 90 % non-condensing |
| Altitude |
120 V ac mains supply voltage up to 5, 000 meters
230 V ac mains supply voltage up to 2,000 meters |
|
| Warranty |
|
Two-year extended warranty (no-cost, available after first-year calibration at the Fluke Biomedical Cleveland Service Center, otherwise standard one year warranty applies) |
|
|
|