The ESA620 Electrical Safety Analyzer represents the next generation in portable, manual electrical-safety testers for the biomedical technician and clinical engineer that are performing tests on medical equipment. With selections of three test loads, protective earth test currents, and two insulation test voltages this versatile product can be used worldwide to test to their standard of choice: IEC60601-1(partial), ANSI/AAMI ES60601, IEC62353, VDE 751, ANSI/AAMI ES1:1993, NFPA-99, AN/NZS 3551, IEC61010.
This multifaceted device performs all primary electrical safety tests, including mains voltage, protective earth resistance, insulation resistance, device current, earth, chassis, and patient leakages as well as several additional leakage tests to comply with EN62353. It comes standard with test leads for 2-wire protective earth measurements, and with optional leads offers the time-saving 4-wire technique. Additionally, the 20-A-device receptacle allows for testing of those devices that need it, so the technician is not limited to only test 15-A equipment. (US only)
New DSP technology offers better accuracy of leakage measurements throughout the ranges specified in the standards, and an intuitive interface guides the user through tests. The ESA620 features a large display to facilitate quick and clear indication of performed tests, criteria associated with test set up, device under test receptacle conditions and results.
Equipped with ten ECG posts, the ESA620 offers simulation of ECG and performance waveforms so both electrical safety and basic tests on patient monitors can be performed with a single connection. The unit is equipped with standard USB interface for computer control and data transfer. When used with optional Ansur computer-based software plug in, the ESA620 allows for standardization of test procedures, as well as result capturing, storage and printing of reports. When combined with Ansur, the ESA620 becomes automated, capturing results and comparing them to standard limits, essentially enabling this manual unit the sophisticated performance of the high end models at half the price.
Key Features
- Compliance with multiple standards
- Three test loads
- Expanded leakage ranges through 10,000 µA
- AC only, DC only and True RMS leakage readings
- 100 % and 110 % mains voltage for mains on applied part (lead isolation) test
- Five different insulation tests
- DSP filter technology
- 20 A equipment current
- More applied parts selections
- ECG and performance waveforms
- Improved user interface
- Easy-to-use applied parts (ECG) connections
- Insulation posts on applied parts connections
- 2- or (optional) 4-wire ground wire resistance
- Large display with adjustable contrast
- Ergonomic design
- Optional Ansur plug-in software
- CE, C-Tick, and CSA for USA and Canada certification
|
| Specifications |
| Available Models |
|
The ESA620 is available in several different models utilizing country-specific or regional-specific test receptacles and power cords |
|
Note: The ESA620 will operate only at a single mains voltage selection: 120 V or 230 V. Dual mains operating voltage is not an ESA620 option |
|
| Voltage Available Test Selections |
|
Mains, Accessible, and Point to Point |
|
| Mains |
| Range(s) |
120 V model: 90 to 132 V ac RMS
220 V model: 180 to 264 ac RMS |
| Measurements |
L1 to L2, L1 to Earth, L2 to Earth |
| Accuracy |
± (2 % of reading + 2 LSD) |
|
| Accessible |
| Range |
0 to 300 V ac RMS |
| Accuracy |
± (2 % of reading + 2 LSD) |
|
| Point to Point |
| Range |
0 to 300 V ac RMS |
| Accuracy |
± (2 % of reading + 2 LSD) |
|
| Earth Resistance |
| Two-Terminal Mode Test Current / Range & Accuracy |
>200 mA ac : 0.0 to 2.0 Ω ± (2 % of reading + 0.015 Ω) |
|
10 to 16 A ac typical : 0.0 to 0.2 Ω ± (2 % of reading + 0.015 Ω)
0.2 to 2.0 Ω ± (5 % of reading + 0.015 Ω) |
| Four-Terminal Mode Test Current / Range & Accuracy |
>200 mA ac : 0.0 to 2.0 Ω ± (2 % of reading + 0.005 Ω) |
|
10 to 16 A ac typical: 0.0 to 0.2 Ω ± (2 % of reading + 0.005 Ω
0.2 to 2.0 Ω ± (5 % of reading + 0.005 Ω) |
| Resistance Tests |
Earth Resistance and Point to Point |
|
| Equipment Current |
| Mode |
AC RMS |
| Range / Accuracy |
0 to 20 A : ± 5 % of reading ± (2 counts or .2A, whichever is greater) |
| Duty Cycle |
15 A to 20 A, 5 min on/5 min off |
|
10 A to 15 A, 7 min on/3 min off |
|
0 A to 10 A continuous |
|
| Leakage Current |
| Modes * |
AC + DC (True RMS) |
|
AC only |
|
DC only |
| Patient load selection (Input Impedance) |
AAMI ES1-1993 Fig.1 |
|
IEC 60601: Fig 15 |
|
IEC 61010: Fig. A-1 |
| Crest factor |
≤ 3 |
| Ranges |
0.0 to 199.9 μA |
|
200 to 1999 μA |
|
2.00 to 10.00 mA |
| Frequency Response/ Accuracy |
DC to 1 kHz : ± (1 % of reading + 1μA or 1 LSD, whichever is greater) |
|
1 kHz to 100 kHz : ± (2 % of reading + 1μA or 1 LSD, whichever is greater) |
|
100 kHz to 1 MHz : ± (5 % of reading + 1μA or 1 LSD, whichever is greater) |
| MAP Voltage |
Additional residual leakage up to 4 µA @ 120 V ac, and 8 µA @ 240 V ac |
| Leakage Tests |
Earth (ground wire) |
|
Chassis (enclosure) |
|
Patient (lead to ground) |
|
Patient Auxiliary (lead to lead) |
|
Mains on Applied Part (lead isolation) |
|
Direct Equipment |
|
Direct Applied Part |
|
Alternative Equipment |
|
Alternative Applied Part |
|
Accessible |
| Mains on Applied Part |
110 % of Mains @ 230 V for IEC 60601 |
| Test Voltage |
100 % of Mains for AAMI, @ 115 V per AAMI |
|
100 % of Mains @ 230 V per 62353 |
|
| Differential Leakage |
| Ranges |
10 to 199 µA |
|
200 to 1999 µA |
|
2.00 to 20.00 mA |
| Accuracy |
± 10 % of reading ± (2 counts or 20 µA, whichever is greater) |
|
| Insulation Resistance |
| Ranges / Accuracy |
0.5 MΩ to 20 MΩ : ± (2 % of reading + 2 counts) |
|
20 MΩ to 100 MΩ : ± (7.5 % of reading + 2 counts) |
| Source Test Voltage |
500 V dc |
|
250 V dc |
| Insulation Resistance Tests |
Mains-PE, AP-PE, Mains- PE, Mains-NE (non-earthed accessible conductive part) and AP- NE (non-earthed accessible conductive part) |
| Maximum Load Capacitance |
1 µF |
|
| ECG Performance Waveforms |
| Accuracy |
± 2 % |
|
± 5 % for amplitude of 2 Hz Square wave only, fixed @ 1 mV Lead II configuration |
| Waveforms / Rates |
ECG Complex (BPM) : 30, 60, 120, 180, and 240 |
| Ventricular Fibrillation |
|
| Square wave (50 % duty cycle) (Hz) |
0.125 and 2 |
| Sine wave (Hz) |
10, 40, 50, 60, and 100 |
| Triangle wave (Hz) |
2 |
| Pulse (63 ms pulse width) |
30 and 60 |
|
| Power Ratings |
| Mains Voltage Outlet |
120 VAC : 230 VAC |
| Mains Voltage Inlet Power Range |
90 to 132 VAC RMS : 180 to 264 VAC RMS |
| Maximum Current |
20 A : 16 A |
| Hz |
50 or 60 : 50 or 60 |
|
| Physical Case |
| Dimensions (cm) |
31 x 23 x 10 |
| Dimensions (inches) |
12.2 x 9 x 2.9 |
| Weight |
4.7kg (10.25 lb) |
|
| Environmental Specifications |
| Operating Temperature |
10 ºC to 40 ºC (50 ºF to 104 ºF) |
| Storage Temperature |
-20 ºC to 60 ºC (-4 ºF to 140 ºF) |
| Operating Humidity |
10 % to 90 % non-condensing |
| Altitude |
To 2,000 meters @ 230 V ac (mains supply voltage) |
|
To 5,000 meters @ 120 V ac (mains supply voltage) |
|
* Modes are available in all leakage tests with the exception of MAP leakages that are available only in True RMS