Fluke Biomedical >> Infusion Device Analyzer >> IDA4 Plus

IDA4 Plus - Infusion Devica Analyzer  

 

  • ภาพรวม
  • ลักษณะเด่น
  • ลักษณะเฉพาะ
  • ดาวน์โหลด
The IDA 4 Plus Multi-Channel Infusion Device Analyzer provides full testing capability for infusion pumps. Available in 1-, 2-, 3-, or 4-channel versions for testing multiple infusion devices simultaneously. (See “Models.”)
  • Tests up to four infusion pumps simultaneously
  • Compatible with virtually any type of infusion device
  • Instantaneous and average flow measurement
  • Occlusion pressure measurements to 45 psi
  • Single- and dual-flow (piggyback) testing
  • Full PCA testing (bolus volume, lockout time, and automated external triggering)
  • Autostart mode enables unit to begin testing only when fluid is detected
  • On-board graphing of pressure and flow
  • Built-in memory to save test results for printing or downloading to computer
  • Hydrograph graphical software to control unit and display results via PC
  • Automated testing through Fluke Biomedical medTester 5000C (US only)
Specifications
Average Flow Rate Technique
 Calculated by measuring a volume over time
Range
 0.5 ml/hr to 1000 ml/hr
Accuracy
 1% of reading + 1 LSD for flows of 16-200 ml/hr for volumes over 20 ml; otherwise, 2% of reading + 1 LSD after delivery of 10 ml
Volume Measurements Technique
 Volume measured directly by the transducer in minimum sample sizes of 60 microliters
Range
 0.06 ml to 9999 ml
Accuracy
 1% of reading + 1 LSD for flows of 16 to 200 ml/hr for volumes over 20 ml; otherwise, 2% of reading ± 1 LSD after delivery of 10 ml
Minimum Bolus Volume
 0.5 ml
Maximum Bolus Count
 99
PCA Measurements Technique
 Volume is measured directly by the transducer in minimum bolus volumes of 0.5 ml. The measurement is made with a continuous rate between 0 and 30 ml/hr. The bolus flow rate should be at least four times the basal flow rate for reliable detection of boluses.
Maximum Bolus Volume
 9999 ml
Accuracy
 See volume measurement
Pressure Measurements Technique
 Direct occlusion of the infusion line and measurement of pressure prior to the glass transducer.
Range
 0 PSI to 45 PSI and equivalents in mmHg and kPa
Accuracy
 1% of full scale + 1 LSD
Back Pressure
 - 100 mmHg to + 300 mmHg
Supply Voltage
 Autoswitching to any voltage from 90 V ac to 260 V ac
Dimensions (LxWxH)
 30.2 cm x 22.9 cm x 18.1 cm
 (11.8 in x 9 in x 7.2 in)
Weight
 4.9 kg (10.8 lb)