Fluke Biomedical >> Defibrillator / External Pacer Analyzer >> Impulse 6000D / 7000DP
  |
Impulse 6000D / 7000DP - Debrillator / External Pacer Analyzer |
|
 |
- ภาพรวม
- ลักษณะเด่น
- ลักษณะเฉพาะ
- ดาวน์โหลด
The Impulse 6000D Defibrillator Analyzer and Impulse 7000DP Defibrillator/Transcutaneous Pacer Analyzer Test Systems are rugged, portable precision test instruments that ensure proper operation and ultimate performance of critical life-support cardiac-resuscitation equipment.
The Impulse 6000D and Impulse 7000DP test capabilities encompass the spectrum of worldwideestablished pulse shapes, showcase breakthrough AED technology compatibility, and outperform in accuracy and standards. Additionally, the Impulse 7000DP incorporates the tests and the extensive range of test loads and measurement algorithms needed to test external transcutaneous pacemakers.
In conjunction with an Impulse 7000DP, the Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω and 200 Ω for defibrillator performance testing. A standard USB interface enables computer control and data transfer, and optional Ansur PC-based automation software increases productivity by outfitting users with an easy-to-use method to standardize testing procedures and capture, print and document data.
Key Features
- Impulse 7010 Defibrillator Selectable Load Accessory provides multiple loads of 25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150 Ω, 175 Ω, and 200 Ω to comply with IEC 60601-2-4 standard (optional)
- Lown, Edmark, trapezoidal, biphasic and pulsed-biphasic defibrillation technology compatibility
- AED technology compatibility
- First-class measurement accuracy ± 1 % of reading + 0.1 J
- Intuitive user interface and backlight, easy-to-read display
- Portable, rugged, easy to carry
- Long-lasting, rechargeable battery
- Pacer brand selections
- Pacer input protected against defibrillator output (7000DP only)
- 10 independent ECG outputs that provide 12 lead combinations for standardized clinical signals
- Flexible heart-rate settings (1 BPM step) facilitate rate meter accuracy and alarm testing
- DSP-based measurements enable future firmware and waveforms upgrade
- Unique integrated posts for secure connections
- Two-year extended warranty (no-cost extended warranty available after first-year calibration at any Fluke Biomedical authorized service center)
- Optional Ansur test automation software to standardize testing procedures, capture waveforms and test results, and print and document test results
- Designed, tested, and built to incomparable Fluke quality standards
|
| General Specifications |
| General Specifications |
| Temperature, operating |
10 °C to 40 °C (50 °F to 104 °F) |
| Temperature, storage |
-20 °C to +60 °C (-4 °F to +140 °F) |
| Humidity |
10 % to 90 % non-condensing |
| Display |
LCD display |
| Communications |
USB device port for computer control |
| Modes of operation |
Manual and remote |
| Power |
Internal rechargeable NiMH battery pack for nine hours (typical) operation after full charge or the battery charger can operate the analyzer and charge the battery simultaneously |
| Battery charger |
100 V to 240 V input, 15 V/1.5 V A output. For best performance, the battery charger should be connected to a properly grounded ac receptacle |
| Enclosure |
ABS plastic housing |
| Dimensions |
32 cm x 24 cm x 13 cm |
|
(13 in x 9.5 in x 5 in) |
| Weight |
3.02 kg (6.6 lb, 0.1 oz) |
| Safety standards |
CE: IEC/EN61010-1 2nd Edition; Polution degree 2 CAN/CSA-C22.2 No. 61010-1; UL61010-1 C-Tick: Australian EMC |
|
Defibrillator Analyzer Specifications
Energy Output Measurement |
| Compatible Defibrillator Waveshapes |
|
Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic |
|
AC Pulsed Biphasic waveform has not been approved in the United States. |
|
| Autoranged Measurement |
|
| Accuracy |
| 0.1 J to 360 J |
± (1 % of reading + 0.1 J) |
| 360 J to 600 J |
± (1 % of reading + 0.1 J), typical |
|
Note: For Pulsed Bi-Phasic defibrillator, specified accuracy is ±(1.5 % of reading + 0.3 J) on both ranges. |
|
| Load resistance |
| Resistance |
50 Ω |
| Accuracy |
±1 %, non-inductive (<2 µH) |
|
| Pulse trigger level |
|
| Pulse width |
| Range |
1 ms to 50 ms |
| Accuracy |
±0.1 ms |
|
| Voltage |
| Range |
20 V to 5000 V |
| Accuracy |
± (1 % of reading + 2 V) |
|
| Current |
| Range |
0.4 A to 100 A |
| Accuracy |
± (1 % of reading + 0.1 A) |
|
| Tilt (biphasic and pulsed biphasic) |
| Range |
1 % to 99 % |
| Accuracy |
± 1 digit |
|
| Interphase delay (biphasic and pulsed biphasic) |
| Range |
0.1 ms to 9.9 ms |
| Accuracy |
± 0.1 ms |
|
| Frequency (pulsed biphasic only) |
| Range |
2000 Hz to 8000 Hz |
| Accuracy |
± 1 % of reading |
|
| Duty cycle (pulsed biphasic only) |
| Range |
1 % to 99 % |
| Accuracy |
± digit |
|
| Sample rate |
|
| Maximum Average Power |
|
12 W, equivalent to 10 defib pulses of 360 J every 5 minutes |
|
| Scope Output |
|
Autorange:2000:1, 400:1 and 80:1: dependant on the range |
|
| Waveform Playback |
|
- Output: BNC
- Output impedance: 50 Ω
- Amplitude Accuracy: ± 5 %
|
|
| Charge Time Measurement |
| Range |
0.1 s to 100 s |
| Accuracy |
± 0.05 s, typical |
|
| Synchronization Test (Elective Cardioversion) |
| Delay Time Measurement |
Timing window: : ECG R-wave peak to the defib pulse peak |
|
Range: -120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380 ms following the R-wave peak |
|
Resolution: 1 ms |
|
Accuracy: ± 1 ms |
| ECG waves |
Normal Sinus Rhythm (NSR): 10 BPM to 180 BPM, in 1 BPM steps |
|
Atrial fibrillation: Coarse and fine |
|
Monomorphic Ventricular Tachycardia: 120 BPM to 240 BPM, in 5 BPM steps |
|
Asystole: Flat line |
|
| Automated Defibrillator Test ECG Waves |
| Normal Sinus |
10 BPM to 300 BPM, in 1 BPM steps |
| Ventricular Fibrillation |
Coarse and fine |
| Monomorphic Ventricular Tachycardia |
120 BPM to 300 BPM, in 5 BPM steps |
| Polymorphic Ventricular Tachycardia |
5 types |
| Asystole |
Flat line |
|
| ECG Waves |
| ECG General |
| Lead configuration |
12-lead simulation. RA, LL, LA, RL, V1-6 with independent signal outputs |
| Lead-to-lead impedance |
1000 Ω (nominal) |
| Rate accuracy |
± 1 % of nominal |
|
| ECG Amplitudes |
| Reference lead |
Selectable, Lead II (default) or Lead 1 |
| Settings |
0.05 mV to 0.45 mV, in 0.05 mV steps
0.5 mV to 5 mV, in 0.5 mV steps |
| Accuracy (All Performance waves and Normal Sinus R waves) |
Lead II: ± 2 %
All other leads: ± 5 %
Defib paddles: ± 5 % |
| Amplitude of ECG signals relative to amplitude setting (in percent) |
Lead II reference: |
|
Performance waves and R wave detection |
|
Lead I – 70 % |
|
Lead II – 100 % |
|
Lead III – 30 % |
|
Lead V1 – 100% |
|
Lead V2 – 100% |
|
Lead V3 – 100 % |
|
Lead V4 – 100 % |
|
Lead V5 – 100 % |
|
Lead V6 – 100 % |
|
Normal Sinus waves |
|
Lead I – 70 % |
|
Lead II – 100 % |
|
Lead III – 30 % |
|
Lead V1 – 24% |
|
Lead V2 – 48% |
|
Lead V3 – 100 % |
|
Lead V4 – 120 % |
|
Lead V5 – 112 % |
|
Lead V6 – 80 % |
|
Lead I reference: |
|
Performance waves and R wave detection |
|
Lead I – 100 % |
|
Lead II – 150 % |
|
Lead III – 50 % |
|
Lead V1 – 100% |
|
Lead V2 – 100% |
|
Lead V3 – 100 % |
|
Lead V4 – 100 % |
|
Lead V5 – 100 % |
|
Lead V6 – 100 % |
|
Normal Sinus waves |
|
Lead I – 100 % |
|
Lead II – 150 % |
|
Lead III – 50 % |
|
Lead V1 – 24% |
|
Lead V2 – 48% |
|
Lead V3 – 100 % |
|
Lead V4 – 120 % |
|
Lead V5 – 112 % |
|
Lead V6 – 80 % |
|
| ECG Normal sinus |
| Rates |
10 BPM to 360 BPM, in 1 BPM steps |
|
| ECG High level output (BNC Jack) |
| Amplitude Range |
0.5 V/mV of reference lead setting |
| Accuracy |
± 5 % |
| Output Impedance: |
50 Ω |
|
| ECG on defibrillator Input Load |
|
Same as the Lead II amplitude but limited to ± 4 mV |
|
| ECG Performance Waves |
| Square wave |
2 Hz and 0.125 Hz |
| Triangular wave |
2 Hz and 2.5 Hz |
| Sine waves |
0.05, 0.5, 5, 10, 40, 50, 60, 100, 150, and 200 Hz |
| Pulse |
30BPM and 60 BPM, 60 ms pulse width |
|
| R-Wave detection |
| Waveform |
Haver-triangle |
| Rates |
30, 60, 80, 120, 200 and 250 BPM |
| Widths |
8, 10, 12 and 20 to 200 (by 10) ms |
| Amplitudes |
0.05 mV to 0.45 mV 0.05 mV hard return |
|
0.5 mV to 5 mV, in 0.5 mV steps |
| Accuracy |
± (1 % setting + 1 ms) |
|
| Noise Immunity |
| Wave |
Sine |
| Line Frequency |
50 Hz or 60 Hz ( ± 0.5 Hz) |
| Amplitude |
Range: 0.0 mV to 10 mV, in 0.5 mV steps |
|
Accuracy: ± 5 % |
|
| Transvenous Pacer Pulse Simulation |
| Widths |
Range: 0.1, 0.2, 0.5, 1, and 2 ms. |
|
Accuracy: ± 5 % of setting |
| Amplitudes |
Range: 0 (off) and ± 2, ± 4, ±6, ±8, ±10, ±12. ±14, ±16, ±18, ±20, ±50, ±100, ±200, ±500, ±700 mV |
|
Accuracy: ± (10 % setting + 0.2 mV) |
| Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent) |
Lead II reference: |
|
Lead I – 67 % |
|
Lead II – 100 % |
|
Lead III – 33 % |
|
Lead V1 – 67% |
|
Lead V2 – 67% |
|
Lead V3 – 67 % |
|
Lead V4 – 67 % |
|
Lead V5 – 67 % |
|
Lead V6 – 67 % |
|
Lead I reference: |
|
Lead I – 100 % |
|
Lead II – 150 % |
|
Lead III – 50 % |
|
Lead V1 – 100% |
|
Lead V2 – 100% |
|
Lead V3 – 100 % |
|
Lead V4 – 100 % |
|
Lead V5 – 100 % |
|
Lead V6 – 100 % |
|
| Arrhythmia selections |
| Pacer Interactive (Transcutaneous Pacer, Impulse 7000DP only) |
Demand: 30 BPM to 360 BPM, in 1 BPM steps
Asynchronous
Non-Capture
Non-Function
Threshold (interactive pacing simulation only): 10 mA to 250 mA in 10 mA steps |
| Supraventricular |
Atrial Fibrillation Coarse
Atrial Fibrillation Fine
Atrial Flutter
Sinus Arrhythmia
Missed Beat
Atrial Tachycardia
Paroxysmal Atrial Tachycardia
Nodal Rhythm
Supraventricular Tachycardia |
| Premature |
Atrial PAC
Nodal PNC
PVC1 Left Ventricle (LV)
PVC1 Left Ventricle (LV) Early
PVC1 Left Ventricle (LV) R on T
PVC2 Right Ventricle (RV)
PVC2 Right Ventricle (RV) Early
PVC2 Right Ventricle (RV) R on T
Multifocal PVCs (LV and RV) |
| Ventricular |
PVCs 6/min
PVCs 12/min
PVCs 24/min
Freq Multifocal
Trigeminy
Bigeminy
Pair PVCs
Run 5 PVCs
Run 11 PVCs
Monomorphic VTach: 120 to 300 (5 BPM steps)
Polymorphic VTach: 5 selections
Ventricular Fibrillation: Coarse and Fine
Asystole |
| Conduction |
1° Block
2° Block Type I
2° Block Type II
3° Block
Right Bundle Branch Block (RBBB)
Left Bundle Branch Block (LBBB) |
| Transvenous Pacemaker (ECG simulation only) |
Atrial 80 BPM
Asynchronous 75 BPM
Demand with frequent sinus beats
Demand with occasional sinus beats
AV Sequential
Non-Capture
Non-Function |
|
| Selectable pacer pulse parameters for transvenous simulation (Atrial & Ventricular channels are independently selectable by the user) |
| Atrial Pacer Pulse (Note: relates only to the TV pacer) |
Width: 0.1, 0.2, 0.5, 1, 2 ms
Polarity: + or -
Amplitude: 0 (off), 2 to 20 (by 2), 50, 100, 200, 500, 700 mV |
| Ventricular Pacer Pulse (Note: relates only to the TV pacer) |
Width: 0.1, 0.2, 0.5, 1, 2 ms
Polarity: + or -
Amplitude: 0 (off), 2 to 20 (by 2), 50, 100, 200, 500, 700 mV |
|
| (BNC Jack) |
| Accuracy |
± 5 % for 2 Hz Square Wave , 50 _ output impedance |
|
| Transcutaneous Pacemaker Analyzer Specifications (7000DP only) |
| Test Load Selections |
| Defibrillator Input |
Fixed load: 50 Ω
Accuracy: ± 2 %, non-inductive (<2 μH)
Power rating: 10 defib pulses of 360 J every 5 minutes |
| Pacemaker Input |
Variable load: 50 Ω to 1500 Ω in 50 Ω steps
Accuracy: ± 2 %, non-inductive (< 2 µH)
Power rating: 5 W (average), 40 W (peak) @ 1000 Ω |
|
| Measurements |
| Manufacturer specific algorithms |
GE Responder (1500 & 1700)
MDE 300 (Medical Data Electronics)
Medtronic ERS/Physio Control LIFEPAK
MRL (Medical Research Laboratory/Welch Allyn)
Philips/Agilent HP
Schiller Medical
ZOLL Medical
(plus a general purpose Default Algorithm selection) |
| Current |
4 to 250 mA, ± (1% of reading + 0.02 mA) |
| Pulse rate |
5 to 800 PPM, ± (0.5% of reading + 0.1 PPM) |
| Pulse width |
1 to 100 ms, ± (0.5% of reading + 0.01 ms) |
| Energy |
1 μJ to 2 J, ± (4% of reading + 10 μJ) |
|
| Demand and Asynchronous Mode Test |
| Input pacer pulse rates |
30 to 200 PPM |
| ECG NSR wave |
Rate:10 to 300 BPM in 1 BPM increments.
Amplitude:1 mV
Underdrive rate:10 BPM minimum
Overdrive rate:300 BPM maximum |
|
| Sensitivity Test |
| Automatic interactive threshold detection |
Compatible pacer rates: 30 to 120 PPM. |
| ECG R Wave |
Waveforms:Square, Triangle, Sine
Width:1 to 19 (by 1) ms. 20 to 95 (by 5) ms. 100 to 300 (by 25) ms
Amplitude:0.05 to 0.95 (by 0.05) mV. 1 to 5 (by 0.5) mV
Accuracy:± 5% of setting |
|
| Refractory Period Tests |
| Paced Refractory Period |
20 to 500 ms |
| Sensed Refractory Period |
15 to 500 ms |
| Accuracy |
± 1 ms |
| Pacer pulse rate |
20 to 200 PPM |
| ECG |
Waveform:Triangle wave
Pulse Width:40 ms
Amplitude:1 mV
|
|
| Impulse 7010 Defibrillator Selectable Load Accessory* |
| General specifications |
| Maximum voltage |
5000 V |
| Maximum continuous power |
12 W. equivalent to 10 defib pulses of 360 J every 5 minutes |
| Inductance |
< 2 µH, @25 Ω |
|
< 3 µH, @50 Ω |
|
< 4 µH, @75 Ω and 100 Ω |
|
< 5 µH, @125 Ω |
|
< 6 µH, @150 Ω |
|
< 7 µH, @175 Ω |
|
< 8 µH, @200 Ω |
| Temperature |
Operating: 10 ºC to 40 ºC (50 ºF to 104 ºF) |
|
Storage: -20 ºC to 60 ºC (-4 ºF to 140 ºF) |
| Humidity |
10 % to 90 % non-condensing |
| Dimensions (WxDxH) |
154 mm x 272 mm x 138.7 mm |
|
(6.07 in x 10.71 in x 5.46 in) |
| Weight (net) |
1.54 kg (3 lb 6.2 oz) |
| Safety class |
Complies with EN61010-1 2nd Edition, Class II product |
| Safety and EMC marks |
CE, CSA, and C-Tick |
| Warranty |
2-year extended warranty (no-cost extended warranty available after first-year calibration at any Fluke Biomedical authorized service center) |
| Calibration intervalq |
1 year |
|
| Electrical specifications (for Load Accessory and Analyzer together) |
| Load settings |
25 Ω, 50 Ω, 75 Ω, 100 Ω, 125 Ω, 150, Ω, 175 Ω, and 200 Ω ± 1 % |
| Accuracy |
Energy (all except pulsed biphasic): 2 % o reading + 0.1 J with 25, 75 Ω through 200 Ω loads, 1 % of reading + 0.1 J with 50 Ω load |
|
Energy (pulsed biphasic): 2.5 % or reading + 0.3 J with 25, 75 Ω through 200 Ω loads, 1.5 % of reading + 0.3 J with 50 Ω load |
|
Voltage: 1 % of reading + 2 V with 25 Ω and 50 Ω loads, 2 % of reading + 2 V with 75 Ω through 200 Ω loads |
|
Current: 2 % or reading + 0.1 A with 25 Ω load, 1 % of reading + 0.1 A with 50 Ω through 200 Ω loads |
|
EditRegion3
|
|