Fluke Biomedical >> Patient Simulator >> PS420
 |
PS420 - Patient Simulator |
|
 |
- ภาพรวม
- ลักษณะเด่น
- ลักษณะเฉพาะ
- ดาวน์โหลด
PS420 Multiparameter Patient Simulator
The PS420 Patient Simulator is a handheld, high-performance simulator for testing patient monitors.
Small enough to fit in a pocket, the handy PS420 features a wide variety of simulation capability, including a full range of ECG, respiration, blood pressure, temperature and cardiac output conditions. The tool includes 12-lead ECG, two-channel blood pressure simulation, 35 arrhythmia selections, pacemaker simulation as well as adult and pediatric normal sinus rhythms.
For convenient use, labeled hot keys on the keypad guide users to the most common settings.
Key Features
- Compact, lightweight, pocketsize
- Labeled hot keys for common settings
- 12-lead ECG
- Respiration and temperature selection
- Two-channel blood pressure simulation
- Optional cardiac output
- Adult and pediatric normal sinus rhythms
- 35 arrhythmia selections
- ECG performance waveforms
- ST segment levels
- ECG artifact
- Pacemaker simulation
- RS232 serial port for computer control
- Battery operated
|
| Specifications |
| ECG |
| Normal Rate: |
80 BPM |
| Selectable Rates: |
30, 40, 60, 80, 100, 120, 140, 160, 180, |
|
200, 220, 240, 260, 280, 300 BPM |
| Accuracy: |
+ 1 % |
| Output Impedance: |
500, 1000, 1500, and 2000 Ω for Leads I, |
|
II, and III |
| ECG Amplitudes: |
0.5 mV, 1 mV, 1.5 mV, and 2 mV |
| Amplitude Accuracy: |
+ 2 % Lead II |
| Waveform Type: |
Adult or Pediatric ECG Waveform |
|
| Performance Waveform |
| Lead II Square Wave: |
|
| Pulse: |
|
30 and 60 BPM, 60 ms pulse width |
|
| Sine Wave: |
|
0.5, 4, 10, 40 50, 60 Hz (1 mV amplitude, lead II) |
|
| Triangle Wave: |
|
| ST Segment Analysis |
| Elevated or Depressed: |
|
0.8 mV to + 0.8 mV in 0.1 mV steps |
|
| Pacemaker |
| Pacer Spike Amplitude: |
|
2, 4, 8, and 10 mV in Lead II |
|
| Accuracy: |
|
| Pacer Spike Duration: |
|
0.1, 0.5, 1, 1.5, and 2 ms |
|
| Accuracy: |
|
| Selections: |
|
Asynchronous Pacemaker |
|
Pacer Non-Function |
|
Pacer Non-Capture |
|
Demand Occasional Sinus |
|
Demand Frequent Sinus |
|
AV Sequential |
|
| Blood Pressure |
| Input/Output Impedance: |
|
| Exciter Input Limit: |
|
| Exciter Input Frequency Range |
|
| Transducer Sensitivity: |
|
| Level Accuracy: |
|
| Static Levels BP1: |
|
- 10, 0, 50, 100, 150, 200, 250 mmHg |
|
| Static Levels BP2: |
|
- 10, 0, 80, 160, 240, 320, 400 mmHg |
|
| Channel Selections: |
|
Arterial 120/80, Channel 1 and 2 |
|
Radial Artery 120/80, Channel 1 and 2 |
|
Left Ventricle 120/00, Channel 1 and 2 |
|
Right Ventricle 25/00, Channel 1 and 2 |
|
Central Venous 15/10, Channel 2 |
|
Pulmonary Artery 25/10, Channel 2 |
|
Pulmonary Wedge 10/2, Channel 2 |
|
Left Atrium 14/4; Automatic Swan/Ganz (every 20 sec) |
|
Manual Swan/Ganz (changes when entry is selected), Channel2 |
|
| Note: |
|
Synchronized with all normal sinus rates. Physiologically track all arrhythmia selection |
|
| Cardiac Output |
| must have optional Cardiac Output Adapter Box p/n 2462200 |
| Catheter Type: |
Baxter Edwards, 10 cc |
| Blood Temperature: |
98.6 ˚F (37 ˚C) |
| CO for 35.6 ˚F (2 ˚C): |
3, 5, 7 l/min |
| CO for 68 ˚F (20 ˚C): |
3, 5, 7 l/min |
| Cal Pulse: |
Of 1 ˚C for 1 sec; of Delta 402 Ω for 4 sec. |
| Computational Constant: |
For 35.6 ˚F (2 ˚C) is 0.561; for 68 ˚F (20 ˚C) is 0.608 |
| Left to Right Shunt: |
35.6 and 68 ˚F (2 and 20 ˚C) |
| Faulty Injectate: |
35.6 and 68 ˚F (2 and 20 ˚C) |
| Accuracy: |
+ 5 % |
| Notes: |
Calibrated or uncalibrated cardiac output waves for 4 different CO values. |
|
| Respiration |
| Baseline Impedance: |
|
500, 1000, 1500, and 2000 Ω, Leads I, II, |
|
and III |
|
| Lead Selections: |
|
| Impedance Variations |
|
| Accuracy: |
|
| Rates: |
|
15, 20, 30, 40, 60, 80, 100, 120, and 0 |
|
BPM for Apnea |
|
| Accuracy: |
|
| Apnea: |
|
12 seconds, 22 seconds, 32 seconds, and |
|
continuous |
|
| Temperature |
| Compatibility: |
|
| Temperature: |
|
86, 95, 98.6, 104, and 107.6 ˚F (30, 35, 37, |
|
40 and 42 ˚C) |
|
| Temperature Simulation Accuracy: |
|
| Arrhythmias |
|
|
Base Rate of 80 BPM |
|
Sinus Arrhythmia |
|
Atrial (PAC)* |
|
Missed Beat* |
|
Atrial Tachycardia |
|
Atrial Flutter |
|
Nodal (PNC)* |
|
Nodal Rhythm |
|
Supraventricular Tachycardia |
|
PVC1 Left Ventricular Focus* |
|
PVC 1 Early, LV Focus * |
|
PVC1 R on T, LV Focus* |
|
PVC2 Right Ventricular Focus* |
|
PVC2 Early, RV Focus* |
|
PVC2 R on T, RV Focus* |
|
Multifocal PVCs* |
|
Atrial Fibrillation Coarse/Fine |
|
PVCs 6/minute |
|
PVCs 12/minute |
|
PVCs 24/minute |
|
Frequent Multifocal PVCs |
|
Bigeminy |
|
Trigeminy |
|
Pair PVCs* |
|
Run 5 PVCs * |
|
Run 11 PVCs* |
|
Ventricular Tachycardia |
|
Ventricular Fibrillation Coarse/Fine |
|
Asystole |
|
Conduction Defects |
|
First Degree |
|
Second Degree |
|
Third Degree |
|
Right Bundle Branch Block |
|
Right Bundle Branch Block |
|
*Will go to NSR ECG @ 80 BPM after completion |
|
| Artifacts |
|
|
50/60 Hz |
|
Muscle |
|
Baseline |
|
Respiration |
|
| General Information |
| Power: |
|
9 V battery/battery eliminator |
|
| Housing: |
|
| Display: |
|
2-line by 16-character LCD with keypad |
|
| RS232: |
|
Bidirectional interface, 9600 Baud |
|
| Dimensions: |
|
6.1 in L x 3.7 in W x 1.3 in H ( 15.6 cm L |
|
x 9.4 cm W x 3.4 cm H) |
|
| Weight: |
|
| Temperature Requirements: |
|
Operating: 59 ˚F to 95 ˚F (15˚C to 35 ˚C) |
|
Storage: 32 ˚F to 122 ˚F (0 ˚C to 50 ˚C) |
|
|
|