| Technical Specifications |
| Generator Output |
| Continuous Operation: |
Continuous measurement of power, current, peak-to-peak voltage (closed load only), and crest factor |
| Single Operation: |
Single measurement after the set delay time of the ESU output of power, current, peak-to-peak voltage (closed load only), and crest factor |
| Power Distribution: |
Automatic measurement of power, current, peak-to-peak voltage (closed load only), and crest factor through a user-selectable load range |
| RF Leakage current: |
Provides connections and load configurations to measure HF leakage from both grounded and isolated equipment |
| RECQM: |
Test the “return electrode control quality monitoring” using the QA-ES internal loads |
|
| Modes of Operation |
|
Manual or remote controlled (via Ansur) |
|
| Measurement |
|
True RMS value of applied waveform |
|
| RMS Bandwidth |
| Instrumentation Only: |
30 Hz to 10 MHz (-3 dB) |
| With loads: |
30 Hz to 2.5 MHz (-3 dB) |
|
| Low Frequency Filter |
|
100 Hz filter to avoid low-frequency disturbance or interference |
|
| Current |
|
| Current Accuracy |
|
20 mA to 2200 mA ± 2 % of reading |
|
| Load Resistance |
|
10 Ω to 2500 Ω in step of 25 Ω (@ dc), 2500 Ω to 5200 Ω in step of 100 Ω (@ dc) |
|
| Additional Fixed Load |
|
200 Ω 400 W for 30 s; max 15 % duty cycle |
|
| Crest Factor |
|
The higher of the two peak voltage measurements is used for computation |
|
| Range |
|
| Foot-Switch Output |
|
The foot switch output can be used to trigger the electrosurgical unit |
|
| Peak-to-Peak Voltage |
| 0 kV to 10 kV (closed load only) accuracy: |
± 10 % |
| Note: |
Measurement is taken between the active and dispersive electrodes with closed load only. |
|
| Oscilloscope Output |
|
5 V/A uncalibrated, 100 mA RF current minimum input |
|
| Ansur QA-ES Plug-In Remote Control |
|
All functions and tests in QA-ES may be performed from the PC. |
|
| User-Programmable Test Sequences |
|
Allows unlimited numbers of test sequences with user-programmable templates and test limits. These tests include power distribution test, output test, HF leakage, and RECQM verification. |
|
| Storage and Recall |
|
Protocol formats and data may be stored, recalled, printed out, or transferred. |
|
| Temperature |
| Operating: |
15 ºC to 35 ºC (59 ºF to 95 ºF) |
| Storage: |
0 ºC to 50 ºC (32 ºF to 122 ºF) |
|
| Display |
|
| Alphanumeric Format |
|
| Graphic Mode |
|
| Display Control |
|
5 F-keys, enter, cancel, control knob, and an encoder |
|
| Data Input/Outputs |
|
Parallel printer port and bidirectional RS232 |
|
| Power |
|
115/230 VAC; 48 to 66 Hz, 35 VA |
|
| Housing |
|
| Dimensions (WxDxH): |
|
34.2 cm x 39.5 cm x 13.2 cm (13.5 in x 15.6 in x 5.2 in ) |
|
| Weight |
|