Electrosurgery-device testing has never been easier than with the RF303RS. User-selectable loads (50 to 750 Ω), signal averaging, battery power and portability make the RF303RS a very versatile tool for determining performance of multiple ESU “makes” and models.
Standard Features
- New Signal Averaging Mode (SAMTM)
- New RS232 port
- Battery powered for completely isolated measurement
- Digital display
- HF-leakage-current measurement
- 4-digit numeric LCD with backlight and power-save mode
|
Optional Feature
- All existing units upgradeable to the NEW RF303RS model
|
| Specifications |
| Modes of Operation |
|
Line powered, battery powered, offline (battery maintenance charge) |
|
| Test Parameters |
|
Power (W), HF/RF current (mA), test load (Ω) |
|
| Test-Load Selection |
|
Monopolar and bipolar |
| Number of Selections: |
15 |
| Range: |
50 Ω to 750 Ω |
| Step Size: |
50 Ω |
| Accuracy: |
± 4 % of selected value at 500 kHz |
|
| Current Measurement (Leakage) |
| Range: |
30 mA to 2500 mA RMS |
| Resolution: |
1 mA |
| Accuracy: |
± 2.5 % of reading or ± 15 mA, whichever is greater |
|
| Power Measurement (Output) |
| Range: |
1 to 400 W |
| Resolution: |
0.1 W |
| Accuracy: |
± 5 % of reading or ± 3 W, whichever is greater |
|
| Frequency Response |
| System Response: |
-3 dB points, 1 kHz to 10 MHz at 300 Ω |
|
| CQM Test |
|
50 Ω to 750 Ω, 50-Ω steps (See “Test-Load Selection” for additional specifications) |
|
| Oscilloscope Output |
|
Transformer coupled output, uncalibrated |
|
Connector Type: BNC |
|
| Battery |
| Type: |
Sealed lead-acid |
| Time Between Recharge: |
Two hours (continuous use) |
| Time to Full Charge: |
Eight hours |
| Number of Cycles: |
200 |
|
| Top Panel Input Connectors |
| Connector Type: |
4 mm-diameter safety sockets |
| Input Voltage Limit: |
10,000 V peak |
| Input CurrentLlimit: |
3 A RMS |
|
| Dimensions |
|
6 in H x 13.25 in W x 11.5 in D |
|
(15.2 cm x 33.7 cm x 29.2 cm) |
|
| Weight |
|