The Index 2 is the most versatile optical simulator for oximeters on the market today. This lightweight, portable tool includes preloaded manufacturers’ R-curves and the ability to define other “makes” for most pulse oximeters.
Motion presets, player mode, transmission level control (TLC), and computer commands boost testing ability. The Index 2 can also be configured to include an optional electrical simulation feature with probe test. Optical and electrical simulations allow technicians to isolate problems quickly. The probe test identifies defective probes with quantitative test results.
Key Features
- Portable
- Tenpreloaded manufacturers’ R-curves
- User-definable “makes” for most other manufacturers
- New R-curves for Masimo, Nonin and Philips Medical Systems (formerly Agilent / H-P) oximeters
- Six downloadable R-curve spaces available
- Simultaneous simulation of motion and arterial-oxygen levels
- Arterial wave-amplitude scale, calibrated in units of perfusion
- Tap/shiver motion simulations to explore the impact of motion
- RS-232 port for computer control
- Physiological finger for complete SpO2 tests
- Electrical simulations with probe testing
|
| Specifications |
| O2 |
| Range: |
35 % to 100 % |
| Resolution: |
1 % |
| Accuracy: |
100 % to 75 %: ± 1 % ± accuracy of the pulse oximeter under test; 74 % to 50 %: ± 3 % ± the accuracy of the pulse oximeter under test; < 50 % unspecified |
| Repeatability: |
Repeatability: ± 1 standard deviation |
|
| Rate |
| Range: |
30 BPM - 250 BPM |
| Resolution: |
1 BPM |
| Accuracy: |
1 % ± 1 BPM |
|
| Pulse Amplitude |
| Range: |
0 to 100 % of nominal pleth amplitude |
| Resolution: |
1 % |
|
| Probe test |
| Continuity/Resistance: |
Pulse amplitude is 20 % of maximum pass-through brightness |
| Test Matrix: |
Measures all combinations of possible interconnections in an XX point matrix |
| Range: |
250 Ω to 150 kΩ |
| Accuracy: |
± 5 % of reading |
|
| LED/Detector voltage test |
| Test Format: |
Measures the voltage drop across Red LED, infrared LED, and the photo detector when the internally generated test signal is applied |
| Test Signal: |
Constant current source @ 1 mA |
| Open Circuit: |
2.5 V max |
| Measurement/Display Range: |
0 V to 4 V |
| Accuracy: |
± 5 % of reading, 0.4 V to 4 V |
|
| Dynamic Test |
| Test Format: |
Photodetector/diode response to both the red and infrared light generated by the probe when pulsed by an internal test signal |
| Test Signal: |
Pulsed between the two LEDs; constant current level @ 1 mA |
| Test Results: |
Nominal range of 0 to 2000 |
|
| Checksum |
|
Sum of all locations in the program chip; for service use only |
|
| General Information |
| Display: |
2-line x 24-character super twist LCD |
| Battery Life: |
At least 4 hours of continuous use |
| Battery Charger: |
Lead acid battery charger only
Input:100 V to 250 V, 50-60 Hz, 0.3 A
Output:+12 V dc, 0.5 A |
| Operating temperature: |
15 ºC to 35 ºC (59 ºF to 95 ºF) |
| Storage temperature: |
0 ºC to 50 ºC (32 ºF to 122 ºF) |
| Humidity |
10 % to 90 % non-condensing |
| Dimensions (WxDxH): |
40.6 cm x 45.7 cm x 22.9 cm
(16 in x 18 in x 9 in) |
| Weight: |
4.5 kg (10 lb) |
|