68/878 หมู่8 ถนนรัตนาธิเบศร์ ตำบลบางกระสอ อำเภอเมือง จังหวัดนนทบุรี 11000
โทร. 0-2965-5585, 0-2965-8889 แฟกซ์ 0-2965-5560
line decor
HOME  :  Products  :  Downloads  :  Maps  :  News  :  Promotion  :  Contacts us  :  
line decor
 

เครื่องมือสอบเทียบมาตรฐานทางการแพทย์
Biomedical Engineering Equipments

เครื่องมือแพทย์
Medical Equipments

เครื่องมือวัดอื่นๆ
Other Measurements

 

Fluke Biomedical >> Pulse Oximeter Simulator >> Index 2

Index 2 - Pulse Oximeter Simulator  

 

  • ภาพรวม
  • ลักษณะเด่น
  • ลักษณะเฉพาะ
  • ดาวน์โหลด
The Index 2 is the most versatile optical simulator for oximeters on the market today. This lightweight, portable tool includes preloaded manufacturers’ R-curves and the ability to define other “makes” for most pulse oximeters.

Motion presets, player mode, transmission level control (TLC), and computer commands boost testing ability. The Index 2 can also be configured to include an optional electrical simulation feature with probe test. Optical and electrical simulations allow technicians to isolate problems quickly. The probe test identifies defective probes with quantitative test results.
Key Features
  • Portable
  • Tenpreloaded manufacturers’ R-curves
  • User-definable “makes” for most other manufacturers
  • New R-curves for Masimo, Nonin and Philips Medical Systems (formerly Agilent / H-P) oximeters
  • Six downloadable R-curve spaces available
  • Simultaneous simulation of motion and arterial-oxygen levels
  • Arterial wave-amplitude scale, calibrated in units of perfusion
  • Tap/shiver motion simulations to explore the impact of motion
  • RS-232 port for computer control
  • Physiological finger for complete SpO2 tests
  • Electrical simulations with probe testing

Specifications
O2
Range:  35 % to 100 %
Resolution:  1 %
Accuracy:  100 % to 75 %: ± 1 % ± accuracy of the pulse oximeter under test; 74 % to 50 %: ± 3 % ± the accuracy of the pulse oximeter under test; < 50 % unspecified
Repeatability:  Repeatability: ± 1 standard deviation
Rate
Range:  30 BPM - 250 BPM
Resolution:  1 BPM
Accuracy:  1 % ± 1 BPM
Pulse Amplitude
Range:  0 to 100 % of nominal pleth amplitude
Resolution:  1 %
Probe test
Continuity/Resistance:  Pulse amplitude is 20 % of maximum pass-through brightness
Test Matrix:  Measures all combinations of possible interconnections in an XX point matrix
Range:  250 Ω to 150 kΩ
Accuracy:  ± 5 % of reading
LED/Detector voltage test
Test Format:  Measures the voltage drop across Red LED, infrared LED, and the photo detector when the internally generated test signal is applied
Test Signal:  Constant current source @ 1 mA
Open Circuit:  2.5 V max
Measurement/Display Range:  0 V to 4 V
Accuracy:  ± 5 % of reading, 0.4 V to 4 V
Dynamic Test
Test Format:  Photodetector/diode response to both the red and infrared light generated by the probe when pulsed by an internal test signal
Test Signal:  Pulsed between the two LEDs; constant current level @ 1 mA
Test Results:  Nominal range of 0 to 2000
Checksum
 Sum of all locations in the program chip; for service use only
General Information
Display:  2-line x 24-character super twist LCD
Battery Life:  At least 4 hours of continuous use
Battery Charger:  Lead acid battery charger only
Input:100 V to 250 V, 50-60 Hz, 0.3 A
Output:+12 V dc, 0.5 A
Operating temperature:  15 ºC to 35 ºC (59 ºF to 95 ºF)
Storage temperature:  0 ºC to 50 ºC (32 ºF to 122 ºF)
Humidity  10 % to 90 % non-condensing
Dimensions (WxDxH):  40.6 cm x 45.7 cm x 22.9 cm
(16 in x 18 in x 9 in)
Weight:  4.5 kg (10 lb)