Fluke Biomedical >> Gas Flow Analyzer >> VT Plus HF

VT Plus HF - Gas Flow Analyzer  
  • ภาพรวม
  • ลักษณะเด่น
  • ลักษณะเฉพาะ
  • ดาวน์โหลด
The VT PLUS HF is Fluke Biomedical’s premier general-purpose gas-flow analyzer. In addition, special display modes and bi-directional flow make it perfect for fully and efficiently testing both conventional mechanical ventilators and high-frequency ventilators. EC.6.20 now requires 100 % completion of scheduled life-support device preventive maintenance every year, and VT PLUS HF can help meet those requirements. Multiple special-function tests make troubleshooting quick and efficient.

VT PLUS HF has the capability to measure either high- or low-flow and pressure, replacing the need for gauges and flow meters. It measures 21 ventilator parameters and can display all of them on one screen. Results can be printed directly from the unit or from a PC with included Windows-compatible software. VT PLUS HF also has onboard graphing capability and shows the minimum, maximum, average, and absolute measurement for all parameters.

Learning to use the VT PLUS HF is simple. Technicians control the unit using the VT PLUS HF user-friendly command system, or, if they’re familiar with the RT-200, they can switch to a special control mode that uses RT-200-style commands.

VT PLUS HF can be operated with a variety of precision test lungs to ensure that ventilators are tested to manufacturers’ specifications and clinical expectations with a fully NIST-traceable testing system.
Key features
  • Bi-directional flow, pressure, volume, and oxygen concentration, and pressure measurements
  • Low- and high-pressure, and flow measurement capability
  • Special HF mode—up to 900 BPM (15 Hz)
  • RS-232 and printer ports
  • Included Windows-compatible graphics software
  • All 21ventilator parameters displayed at once on one screen
  • Operation by user-friendly VT PLUS HF command mode or special RT-200 command mode
  • Minimum, maximum, average, absolute, and graph for all parameters
  • Multiple special-function tests for efficient troubleshooting

Optional Feature

  • Operation with a variety of precision test lungs available from Fluke Biomedical to complete a fully NIST-traceable ventilator testing system
Specifications
Power
 100 V ac to 240 V ac, 50/60 Hz
Maximum over-voltage:  264 Vac
Display
 320 x 240 LCD with CFL backlight
Viewing Area:  10.1 cm x 8.2 cm (3 in x 4 in)
 Blue on white background
Operational Modes
 Manual mode, for simple tests or troubleshooting
 Computer-control mode, using RS-232 serial port for special applications
 Use of VT PLUS HF with VT Plus for Windows, for recording graphs and logging data to a computer
Oxygen Measurement
Range:  0 % to 100 %
Accuracy:  ±2 % FSO
Resolution:  0.1 % O2
Transducer Location:  Internal
Gas Compatibility
 Air, O2, CO2, N2, N2O, He, mixtures or user-defined
Gas Reference Units
 ATP, STPD0, STPD21 and BTPS
Output Ports
 RS-232 serial port and parallel printer port
Test Parameters
Continuous Flow:  Low Flow: ± (2 % of reading and 1 % of range)
High Flow: ± (2 % of reading and 1 % of range)
Volumetric Flow:  
 Low-Flow:
 Flow Range: -25 lpm to 25 lmp
 Accuracy: ± 2 % of reading or ± 1 % of range, whichever is greater
 Frequency Response: > 25 Hz or t10-90 < 40 ms, whichever is greater
 Low Flow Dropout: 0.01 lpm
 Breath-Defect Threshold: 0.5 lpm
 Maximum-Flow Rate: 50 lpm
 Volume Range: > ± 60 L
 Sample Rate: 100 Hz
 Resolution: 0.01 lpm flow > 1 lpm; 0.001 lpm flow < 1 lpm
 Dynamic resistance: < 2.5 cmH2O @ 5 lpm
 Fittings: 15 mm OD, 1:40 conical male; 0.25 in NPT ID per ASTM F-1054 aluminum with black anodized finish
 Notes:
  • Tidal-volume accuracy: ± 3 % of reading or ± 2 ml, whichever is greater
  • Volume accuracy tested to 1 liter
  • Flow accuracy is specified for dry air or oxygen
  • Below 3.0 lpm, measurement accuracy is obtained by allowing the VT PLUS HF to fully warm up or manually zeroing before reading or documenting measurement
 High-Flow:
 Flow Range: -300 lpm to 300 lmp
 Accuracy: ± 2 % of reading or ± 2 % of range, whichever is greater
 Frequency Response: > 25 Hz
 High-Flow Dropout: 25 lpm
 Breath-Detect Threshold: 2 lpm
 Maximum-Flow Rate: 500 lpm
 Volume Range: > ± 60 L
 Dynamic Resistance: < 2 cmH2O @ 60 lpm
 Sample Rate: 100 Hz
 Resolution: 0.01 lpm
 Fittings: 22 mm OD, 1:40 conical male; 15 mm ID, 1:40 conical female per ASTM F-1054 aluminum with black anodized finish
 Notes:
  • Tidal-volume accuracy: ± # % of reading or ± 10 ml, whichever is greater
  • Volume accuracy tested to 7 liters
  • Flow accuracy is specified for dry air or oxygen
 Low-Pressure:
 Range: ± 500 mmHg (10 psi)
 Accuracy: ± 0.8 % of reading or ± 1.5 mmHg, whichever is greater
 Frequency Response: > 10 Hz
 Resolution: 0.1 mmHg
 Fittings: Luer lock, stainless steel
 Maximum Applied Pressure: 60 psi
 Sample Rate: 100 Hz
 Operating Pressure: 30 psi
 Note:
Fluid pressure may be applied to the positive port; however, fluids should be kept from entering the pressure port by using a suitable length of connection tubing
 
High-Pressure:
 Maximum Applied Pressure: 150 psi
 Range: ± 100 psi
 Accuracy: ± 1 % of reading or ± 0.3 psig, whichever is greater
 Frequency Response: > 10 Hz
 Resolution: 0.1 psi
 Sample Rate: 100 Hz
 Fittings: DISS connector, stainless steel

 Airway-Pressure:
 Maximum Applied Pressure: 20 psi
 Range: ± 120 cmH2O
 Accuracy: ± 0.75 % of reading or ± 0.5 cmH2O, whichever is greater
 Frequency Response: > 25 Hz or t10-90 < 40 ms, whichever is greater
 Resolution: 0.1 cmH2O
 Fittings: Internally connected at the transducer distal end
 Note:
Airway pressure is internally tapped off the proximal-flow sensor port, which is the port closest to the exhaust port on the VT PLUS HF
Ventilator Parameter
Inspiratory and expiratory tidal volume  Range:As specified in high-flow/low-flow spec
 Accuracy: As specified in high-flow/low-flow spec
Expiratory minute volume  Range: 0 L to 60 L
 Accuracy: ± 3 of % or 250 ml
Breath rate  Range: 0.5 BPM to 150.0 BPM
 Accuracy:± 1 of %
Inspiratory-to-expiratory time ratio (I:E ratio)  Range: 1:200 to 200:1
 Accuracy: ± 2 % or ± 0.1 s
Inspiratory time  Range: 0 s to 60 s
 Accuracy: ± 1 % or ± 0.1 s
Expiratory time  Range: 0 s to 90 s
 Accuracy: ± 1 % or ± 0.1 s
Peak inspiratory pressure  Range: ± 120 cmH2O
 Accuracy: ± 3 % or ± 1 cmH2O
Inspiratory pause pressure  Range: ± 120 cmH2O
 Accuracy: ± 3 % or ± 1 cmH2O
Mean airway pressure  Range: ± 80 cmH2O
 Accuracy: ± 3 % or ± 0.5 cmH2O
Positive end-expiratory pressure (PEEP)  Range: -5 to 40 cmH2O
 Accuracy: ± 3 % or ± 0.5 cmH2O
Lung compliance  Range: 0 to 150 ml/cmH2O
 Accuracy: ± 5 % or ± 5 ml/cmH2O
Inspiratory hold time  Range: 0 s to 60 s
 Accuracy: ± 1 % or ± 0.1 s
Expiratory hold time  Range: 0 s to 90 s
 Accuracy: ± 1 % or ± 0.1 s
Peak expiratory flow  Range: 0 lpm to 300 lpm
 Accuracy: ± 3 % or ± 2 lpm
Peak inspiratory flow  Range: 0 lpm to 300 lpm
 Accuracy: ± 3 % or ± 2 lpm
Flow bias  Range: 0 lpm to 30 lpm
 Accuracy: ± 2 % or ± 0.5 lpm
Operating Environment Conditions
Temperature Range:  10 °C to 40 °C
Ambient Humidity:  0 % to 80 % noncondensing to 31 °C, decreasing to 50 % at 40 °C
Barometric Pressure:  8 psig to 18 psig
Storage Environment Conditions
Temperature Range:  -25 °C to -50 °C
Humidity:  0 to 95 % noncondensing
Dimensions (WxDxH)
 25.4 cm x 25.4 cm x 12.7 cm
 (10 in x 10 in x 5 in)
Weight
 4.53 kg (10 lb)